1.SPO2 AND TEMP Pulse Oximeter(KCKP012)

SPO2 AND TEMP Pulse Oximeter

Parameters:

Three Parameters:SPO2,TEMP,Pulse Rate.

Waveforms:

Waveform: SPO2 waveform-PLETH

Features:

1.Compact, small, light, easy for carrying and handling.

2.LCD display of measurement and PLETH.

3.Intelligent parameter monitoring interface.

4.Audio and visual alarm.

5.Storage and review of patient's trend data up to 36 hours which is never lost even when power's off.

6.Common AA alkaline or rechargeable batteries to support 7 continuous working hours; 

7.Battery capacity indicator.

8.Automatic power-off

9.Patient's trend can be transmitted to a personal computer for display, saving and printing.

10.Suitable for adult, pediatric, infant and neonate patients.

H.S.Code: 90189090

Dimension:130x 72x 34mmNet Weight: 145gGross Weight: 650gPackage Size: 200x150x70mm

Color: Blue

Standard accessories

one SpO2 adult finger probe

one TEMP Skin face probe

one communication cable

one software CD 

Specifications:

Patient Range

 Adult, pediatric, infant, neonate patients

Oxygen saturation (SpO2)

SpO2

Measurement range: 35%~100% 

Accuracy: ±2%(70%~100%) ,0%~69% unspecified

Resolution: 1%  

Pulse Rate

Measurement range:30 bpm -250 bpm  

Accuracy:±2bpm  

Resolution:1bpm  

Temperature (TEMP)

Channel: 1

Input: Body surface thermal-sensitive resistor temperature sensor

Measurement range:0 oC~50°C  

Accuracy:±0.2°C 

Resolution:0.1°C  

Alarm

Mode:Audio and visual alarm  

Setup:User-adjustable high and Low limits 

Storage and Review

36 hours SpO2 \PR\TEMP trend data with corresponding date and time   

Display Mode

LCD screen:128×64 dot matrix homochromy LCD  

                  Measurements and information display

LED indicator:Red and green Dual-color LED 

                  Red LED: insufficient battery power indicator 

Green LED: pulsating rhythm indicator. 

Operation Environment

Temperature:

Working:0 oC ~ 45oC  

Transport and storage:-20 oC ~ 60 oC  

Humidity:     

Working:15%~ 95%(non-condensing)  

Transport and storage:10 %~ 95%(non-condensing)   

Altitude: 

Working:86 kPa ~ 106 kPa 

Transport and storage:50 kPa ~ 106 kPa

Power Requirements

Battery type:common 1.5V AA alkaline or rechargeable batteries